ORIGINAL ARTICLE
Year : 2020  |  Volume : 1  |  Issue : 1  |  Page : 38-42

Installation, commissioning, and performance evaluation of Bhabhatron-II TAW – An Indian-made telecobalt unit


Department of Radiation Oncology, Dr. B. Borooah Cancer Institute, Guwahati, Assam, India

Date of Submission11-Jun-2020
Date of Decision24-Jun-2020
Date of Acceptance29-Jun-2020
Date of Web Publication25-Sep-2020

Correspondence Address:
Mr. Shachindra Goswami
Department of Radiation Oncology, Dr. B. Borooah Cancer Institute, Guwahati, Assam
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/JRMT.JRMT_12_20

Rights and Permissions
  Abstract 


Aim: The aim of this study was to report commissioning and performance evaluation of Bhabhatron-II TAW (an Indian made machine). Materials and Methods: Commissioning comprises of following quality assurance (QA) (a) electrical QA, (b) mechanical QA, (c) radiation QA and (d) survey of installation. Electrical QA were done for various components of the unit for its functionality. Mechanical QA were done and the measured data were verified as per guidelines of Atomic Energy Regulatory Board (AERB), India and vendor's specified values. Cylindrical ionization chamber, electrometer, ready pack films, water phantom, thermometer and barometer were used during radiation QA and survey of installation procedures. Moreover, the data so generated were fed to treatment planning system (TPS) for dose calculation and also kept in proper tabular manner for clinical use. Results: Couch and collimator parameters were found to comply the tolerance criteria. Source strength of 172.2 roentgen per minute at 1 meter (RMM) was measured with an absolute dose rate of 225.76 cGy/min at dmax for 10cm x 10cm field size at normal treatment distance (NTD). Shutter timer error of 0.004 min was measured and incorporated in the clinical data. A maximum of 3.66 % variation was noted for 300 (15W × 20) physical wedge transmission factor between measured and vendor specified value. Radiation leakage through beam limiting device (BLD) and outside BLD were found to be within the limits prescribed by AERB. Conclusion: Commissioning is a prerequisite process after installation of a radiation therapy machine for medical purpose. Bhabhatron-II machine has several advantages over other telecobalt machine. Features like high source storage capacity of machine head, one pair of asymmetric jaw, closure of jaws to 0 cm x 0 cm during radiation emergency and presence of one motorized wedge filter makes it unique and yet cost effective telecobalt machine..

Keywords: Bhabhatron-II TAW, commissioning, performance evaluation, quality assurance


How to cite this article:
Sharma SB, Sarma G, Barthakur M, Goswami P, Yadav B, Goswami S. Installation, commissioning, and performance evaluation of Bhabhatron-II TAW – An Indian-made telecobalt unit. J Radiat Med Trop 2020;1:38-42

How to cite this URL:
Sharma SB, Sarma G, Barthakur M, Goswami P, Yadav B, Goswami S. Installation, commissioning, and performance evaluation of Bhabhatron-II TAW – An Indian-made telecobalt unit. J Radiat Med Trop [serial online] 2020 [cited 2023 Mar 26];1:38-42. Available from: http://www.jrmt.org/text.asp?2020/1/1/38/296108




  Introduction Top


Cancer is a major public health problem and the incidence is increasing worldwide. About 65% of all cancer deaths globally occurred in low- and middle-income countries (LMICs) in 2012.[1],[2] This is projected to increase to 75% by 2030.[3],[4] An estimated 18.1 million new cancer diagnoses and 9.6 million cancer deaths happened in 2018.[1],[5] As per the WHO estimation, this will increase to 29.5 million new cancer diagnoses and 16.5 million cancer-related deaths annually by 2040.[6] In high-income country like the United States, 1,806,590 new cancer cases and 606,520 cancer deaths are projected to occur in 2020.[7] Patients with cancer in LMICs have a poorer prognosis because of relatively low awareness, late diagnosis, and the lack of access to affordable curative services.[6],[8] More than 50% of all cancer patients require radiation therapy during their course of illness.[9],[10] The main hurdle to improve access to cancer care in LMICs is the high cost of modern radiotherapy machines. Among all the external-beam teletherapy machines available, telecobalt machines are the cheapest. Although telecobalt machines have some shortcomings, they have the advantages of (i) low cost, (ii) low maintenance cost, (iii) lower power need (iv), less machine downtime, and (v) constancy of beam quality. They are suitable for treatments of commonly encountered type of cancers in LMICs.

Before using any radiation facility for medical purpose, every machine needs to undergo commissioning as a prerequisite procedure. Commissioning is the process during which structures, systems, and components of a nuclear and radiation facility, on being constructed, are made functional and verified to be in accordance with design specifications and to have met the performance criteria.[11]

Bhabhatron-II Tungsten, Asymmetric, and Motorized Wedge (TAW) is an Indian-made telecobalt machine. It was designed and manufactured by Bhabha Atomic Research Centre, India. It complies with the radiological safety standards of the International Electrotechnical Commission (IEC).[12] Many cancer centers of LMICs have recently installed Bhabhatron-II TAW for cancer treatment. In this study, we will discuss the various parameters related to commissioning and performance evaluation of Bhabhatron-II TAW installed at Dr. Bhubaneswar Borooah Cancer Institute, Guwahati, a tertiary cancer institute of India. When comparing with other telecobalt machines, Bhabhatron-II TAW machine has some unique features like high-source loading capacity of 250 Roentgen per minute at 1 m (RMM) in the machine head, one pair of asymmetric Y-jaw as Y1 and Y2, capability of automatically closing of jaws to 0 cm × 0 cm in case of source stuck, and one motorized wedge of 60° inside the collimator head of the machine.[13]

This study was approved by the expedited review board of the Institutional Ethics Committee.


  Materials and Methods Top


During installation and commissioning of the Bhabhatron-II TAW telecobalt machine, a series of Quality Assurance (QA) checks were performed in compliance with Atomic Energy Regulatory Board (AERB), India. The QA involves (a) electrical QAs, (b) mechanical QAs, (c) radiation QAs, and (d) survey of installation. Post commissioning and acceptance tests, geometrical and beam data were fed to the treatment planning system for dose calculation. These data were preserved in proper tabular format as a baseline for future quality control (QC) program. Future QC programs were planned as per the literature published by the American Association of Physicist in Medicine Report No. 46.[14]

Electrical QAs are performed with visual and functional observations. In mechanical QAs, the parameters of optical field size verifications, mechanical isocenter with respect to the rotation of gantry and collimators, accuracy of alignment lasers, optical distance indicator (ODI) accuracy, movements of couch (treatment table) in various directions, collimator and gantry angle accuracy, parallelism and orthogonality of jaws, etc., were checked for their specified tolerance prescribed by AERB, India.

Radiation QA involves the measurements of shutter timer error, optical and radiation field congruence, isocenter accuracy, absolute radiation output of the unit, etc., For carrying these QA tests, the following equipments are used:

  • Cylindrical ion chamber of 0.6 cc volume (make: PTW) with electrometer
  • Water phantom (30 cm × 30 cm × 30 cm) with a waterproof sleeve
  • Calibrated thermometer and barometer
  • Ionization-based survey meter (Fluke Biomedical System, Model: 451)
  • Radiochromic ready-packed verification film (Kodak X-OMAT V Film)
  • Slab phantom of 1-cm thick slabs.


Optical field size verification

Optical field size was verified for field sizes 5 cm × 5 cm, 10 cm × 10 cm, 15 cm × 15 cm, 20 cm × 20 cm, 25 cm × 25 cm, 30 cm × 30 cm, and 35 cm × 35 cm using a graph sheet at 80 cm source to surface distance (SSD) keeping and gantry and collimator 0°. Deviations were noted for the field sizes.

Tolerance: For ≤10 cm × 10 cm: ±1 mm and for >10 cm × 10 cm: ±2 mm

Mechanical Isocenter check (with respect to collimator rotation)

Mechanical isocenter check was performed by placing a graph sheet on the flat couch at 80 cm SSD with gantry 0°. Optical crosshair was initially matched with a point on the graph sheet and then the collimator was rotated in clockwise and counterclockwise direction with its extreme limits. The deviation of the crosshair was observed throughout the procedure and the maximum deviation was noted.

Tolerance: Within 4-mm diameter circle.

Mechanical Isocenter check (with respect to gantry rotation)

A pin with a sharp tip was attached on the couch edge and its tip was matched with isocenter. Now, the gantry was rotated ± 180° and deviation was noted throughout the 360° rotation of gantry.

Tolerance: Within 4-mm diameter sphere.

Laser alignment test

Laser alignment test tool was placed on the couch and was aligned with help of crosshair and ODI. Deviation of the three lasers (sagittal, coronal, and transverse) was checked with the test tool.

Tolerance: ±1.5 mm.

Roentgen per minute at one meter measurement

The 0.6 cc ionization chamber with the buildup cap was placed at isocenter with its long axis parallel to the gantry axis of rotation. The ionization chamber was exposed for 1 min for a field size of 20 cm × 20 cm and electrometer reading was noted.

Tolerance: ±10% from quoted value.

Congruence between Optical and Radiation field

For this test, a 30 cm × 30 cm ready-packed film pinpricked at the optical field (10 cm × 10 cm) borders was placed on the couch at 80 cm SSD. After putting sufficient buildup, the film was exposed to the optically defined field sizes to get an optimal optical density of one. After developing the film, the deviation between the optical and radiation field was noted for all the edges. This procedure can be repeated for different field sizes.

Tolerance: For ≤10 cm × 10 cm: ±1 mm and for >10 cm × 10 cm: ±2 mm

Shutter timer error measurement

For the measurement of shutter timer error, the experimental setup was done as per the International Atomic Energy Agency (IAEA) Technical Report Series (TRS)-398[15] document. Electrometer reading R1 for T = 2 min of exposure was noted and then meter reading R2 for exposure time of T/2 + T/2 min was noted without resetting the electrometer. From the collected readings, shutter timer error was calculated using the following formula.



Output calibration for different field sizes

The measurements of the absolute absorbed dose for various field sizes were carried out as per the IAEA TRS-398 document. Output factors were calculated for various field sizes ranging from 5 cm × 5 cm to 30 cm × 30 cm.



Wedge (physical) factor

After output measurement, the same setup was used for the measurement of wedge transmission factor. Here, the electrometer readings were noted for 1 min of exposure for both the open field and wedge fields. The ratio of the electrometer reading with wedge and without wedge gives the wedge transmission factor for that particular wedge filter. Measurements were taken for 10 cm × 10 cm and maximum wedge fields.

Leakage measurement

During the ON and OFF position of the source, radiation leakage was measured as per the CODE NO. AERB/RF-MED/SC-1 (Rev. 1).[11]

Radiation protection survey

Radiation protection survey was performed for maximum open-field size all around the installation and control console using an ionization-based survey meter for four gantry angles (0°, 90°, 180°, and 270°).


  Results Top


Upon successful installation of the unit, all the required QAs were done for obtaining a license for clinical use.

Electrical quality assurance tests

Tests such as electrical interlocks, gantry head display, control console display, and control console functions were performed for the unit and results were satisfactory.

Mechanical quality assurance tests

Couch and collimator

All mechanical parameters for couch and collimator of the unit were performed and found to be within the specified limits of national competent authority [Table 1].
Table 1: Couch and collimator parameters

Click here to view


Radiation quality assurance tests

Roentgen per minute at one meter

RMM for the unit was found to be 172.2 RMM with an acceptable deviation of 7.62% from the quoted RMM from Board of Radioisotope and Technology, India.

Shutter timer error

It was 0.004 min [Table 2] which needs to be incorporated in the treatment time calculation.
Table 2: Shutter timer error calculation

Click here to view


Output measurement

The output (dose rate) for 10 cm × 10 cm was 225.76 cGy/min [Table 3]. The output of 10 cm × 10 cm was used as a baseline value for subsequent monthly measurements.
Table 3: Absolute absorbed dose measurement as per the International Atomic Energy Agency Technical Report Series-398

Click here to view


Optical and radiation field congruence

Congruence between optical and radiation field size for 5 cm × 5 cm and 10 cm × 10 cm was found to be ≤ 1 mm. For field sizes 15 cm × 15 cm and 20 cm × 20 cm, the deviation was ≤ 2 mm.

Wedge factor

The wedge transmission factor (WF) for various wedges was determined and is tabulated in [Table 4]. A maximum variation of 3.66% between manufacturer specified value and measured WF was noted.
Table 4: Wedge factor measurement (physical)

Click here to view


Leakage measurement

Maximum leakage radiation at 5 cm from source storage surface was found to be 1.5 mR/h (tolerance: ≤20 mR/h) and at 1 m from the source was 0.3 mR/h (tolerance: ≤2 mR/h) in source OFF condition. The maximum transmission through beam-limiting device in source ON condition was measured to be 1.3% (tolerance: 2%). Leakage radiation from the source housing in ON condition in the patient plane was found to be a maximum of 0.12% and average of 0.098% (tolerance: maximum 0.2%, average 0.1%). Leakage radiation from the source housing in ON condition at 1 m from the source in other than the patient plane was found to be 0.3% (tolerance: 0.5%). The maximum radiation during radiation protection survey around the bunker was found to be 1.3 mR/h which was a no occupancy area.

Proper radiation warning symbols were also placed at the maze entrance door and outside the treatment room. Gamma zone monitor was installed inside the bunker.


  Discussion Top


Commissioning and performance evaluation are done to ensure radiation safety and accurate dose delivery of the machine. The guidelines for commissioning and performance evaluation can be obtained from IEC documents or respective national competent authority. However, data such as physical wedge filter transmission factor, couch transmission factor, and lead shielding tray transmission factor which are specific to the machine are provided by the vendor.

Our results were well within the acceptable criteria laid by AERB, India, and it also matched with the specifications provided by the vendor.

Some of the beam data such as percentage depth dose, tissue-to-air ratio, and Tissue-maximum ratio were taken from the published literature for dose calculations during clinical use.[16],[17]

Six of seven physical wedge filter transmission factors have values ≤3% and maximum deviation was found to be 3.66% from vendor specifications. Akinlade et al. found that six wedges of seven wedge filters transmission factor values are ≤3% and maximum deviation of 6%.[18] Mohib-UlHaq et al. reported that four wedges of seven wedge filters have transmission factor values ≤3% and maximum deviation of 5.01%.[19] Tray transmission factors of shielding trays (plain and slotted) were measured and recorded as a baseline value for consistency check in future.

Bhabhatron-II TAW has a maximum source loading capacity which is 250 RMM. The source strength of our machine was 172.2 RMM. The reduction in source strength was due to the unavailability of high RMM source. Akinlade et al. reported loading source strength of 151.3 RMM.


  Conclusion Top


Commissioning and performance evaluation are the prerequisites before starting clinical treatment. It is necessary to perform the commissioning and performance evaluation of the machine according to the guidelines laid by IEC or national competent authority. Bhabhatron-II TAW has many advantages in comparison to earlier telecobalt machines. Features such as high-source loading capacity of machine head (250 RMM), one pair of asymmetric jaw, closer of jaws to 0 cm × 0 cm, and one motorized wedge filter make Bhabhatron-II TAW a unique yet cost-effective telecobalt machine for LMICs.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Global Cancer Observatory; 2020. Available from: https://gco.iarc.fr/. [Last accessed on 2020 Jun 11].  Back to cited text no. 1
    
2.
Parkin DM, Bray F, Ferlay J, Pisani P. Global cancer statistics, 2002. CA Cancer J Clin 2005;55:74-108.  Back to cited text no. 2
    
3.
Cancer Tomorrow; 2020. Available from: https://gco.iarc.fr/tomorrow/home. [Last accessed on 2020 Jun 11].  Back to cited text no. 3
    
4.
The Lancet. GLOBOCAN 2018: Counting the toll of cancer. Lancet 2018;392:985.  Back to cited text no. 4
    
5.
Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin 2018;68:394-424.  Back to cited text no. 5
    
6.
Sivaram S, Majumdar G, Perin D, Nessa A, Broeders M, Lynge E, et al. Population-based cancer screening programmes in low-income and middle-income countries: regional consultation of the International Cancer Screening Network in India. Lancet Oncol 2018;19:e113-2.  Back to cited text no. 6
    
7.
Siegel RL, Miller KD, Jemal A. Cancer statistics, 2020. CA Cancer J Clin 2020;70:7-30.  Back to cited text no. 7
    
8.
Chalkidou K, Marquez P, Dhillon PK, Teerawattananon Y, Anothaisintawee T, Gadelha CA, et al. Evidence-informed frameworks for cost-effective cancer care and prevention in low, middle, and high-income countries. Lancet Oncol 2014;15:e119-31.  Back to cited text no. 8
    
9.
Delaney G, Jacob S, Featherstone C, Barton M. The role of radiotherapy in cancer treatment: estimating optimal utilization from a review of evidence-based clinical guidelines. Cancer 2005;104:1129-37.  Back to cited text no. 9
    
10.
Begg AC, Stewart FA, Vens C. Strategies to improve radiotherapy with targeted drugs. Nat Rev Cancer 2011;11:239-53.  Back to cited text no. 10
    
11.
CODE NO. AERB/RF-MED/SC-1 (Rev. 1): Radiation therapy sources, equipment and installations, Atomic Energy Regulatory Board, India; 2011.  Back to cited text no. 11
    
12.
Sahani G, Kumar M, Dash Sharma PK, Sharma DN, Chhokra K, Mishra B, et al. Compliance of Bhabhatron-II telecobalt unit with IEC standard-Radiation safety. J Appl Clin Med Phys 2009;10:2963.  Back to cited text no. 12
    
13.
Jayarajan K, Kar DC, Sahu R, Radke M, Singh GM. BARC develops Cobalt-60 teletherapy machine for cancer treatment. BARC Newsletter Bhabha Atomic Res Centre 2005;253;10-4.  Back to cited text no. 13
    
14.
Kutcher GJ, Coia L, Gillin M, Hanson WF, Leibel S, Morton RJ, et al. Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40. Med Phys 1994;21:581-618.  Back to cited text no. 14
    
15.
Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. Vienna: Technical Report Series No. 398, International Atomic Energy Agency; 2000.  Back to cited text no. 15
    
16.
Aird EG, Burns JE, Day MJ, Duane S, Jordan TJ, Kacperek A, et al. Central axis depth dose data for use in radiotherapy. Brit J Radiol 1996;25:Suppl 25.  Back to cited text no. 16
    
17.
Khan FM. The Physics of Radiatio Therapy. Appendix A-15. 3rd ed. Philadelphia, USA: Lippincott Williams & Wilkins; 2003.  Back to cited text no. 17
    
18.
Akinlade B, Oyekunle E, Adenipekun A, Elumelu-Kupoluyi T, Folasire A. Dosimetric verification of bhabhatron-II telecobalt unit at the University College Hospital, Ibadan, Nigeria. Cancer Biol 2014;4:22-9.  Back to cited text no. 18
    
19.
Mohib-UlHaq M, Rehman M, Ahmad A, Khan N, Ayub M, Ahmad S. Installation, commissioning and quality assurance tests of first Indian made Tele-therapy cobalt machine (Bhabhatron- II) at Sher-I-Kashmir institute of medical sciences, Srinagar, Jammu and Kashmir. J K Practitioner 2014;19:93-9.  Back to cited text no. 19
    



 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Materials and Me...
Results
Discussion
Conclusion
References
Article Tables

 Article Access Statistics
    Viewed3325    
    Printed171    
    Emailed0    
    PDF Downloaded189    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]